TH120-E Clean Industrial Wall-Mounted Temperature and Humidity Transmitter

The clean rooms and laboratories in the biopharmaceutical industry need to strictly control the ambient temperature and humidity, and require the structure of the on-site installation instrument to prevent dust and ash. TH120-E adopts the wall-mounted installation structure of dust and ash prevention, and the signal cable is connected through the external terminal to facilitate equipment maintenance. It is suitable for the temperature and humidity monitoring of the biopharmaceutical EMS system. For the O3 or VHP sterile environment, the temperature and humidity probe can be equipped with H202 environment-specific filter PF04. The microholes of the filter are filled with MnO2 composite catalyst with the particle diameter of about 10um; When in touch with the filter, VHP or O3 will be decomposed into H2O and O2 to achieve the purpose of protecting the probe and humidity-sensitive element.

    In order to facilitate field equipment maintenance, the TH120 is also equipped with interchangeable digital probes and digital reference probes for analog output channel accuracy verification. There are two reference probes, one of which has a fixed output of 90%RH/5℃, and the other has a fixed output of 10%RH/45℃. In addition, the probe, extension probe rod and signal cable can also be equipped with thread combination structure, which is convenient for regular maintenance and calibration of the probe.

Technical Features

Output signal: two ways 4~20mA or RS485

Shell material: stainless steel panel + flame-retardant PC housing

Calibration interface: Single-point or two-point calibration via calibration button

Digital probe: interchangeable digital probe for easy on-site maintenance

Reference probe: easy to verify the analog output channel on site

External wiring: facilitate the on-site maintenance of the instruments


Technical Data

Input and output

Operating voltage

24V DC ±20%

Current output

Two ways 4... 20mA  RL < 500 Ohm

Power consumption

<20mA (current output)

Channel precision: ±0.2%FS

< 6mA (digital output)

Digital output

RS485 (Modbus RTU)

Measurement Parameter

Relative humidity

  

Temperature

Mearsurement range 

... 100 %RH

Mearsurement range 

−40 … +80°C

Measurement precision 

±1%RH (0 … 90%RH & 20℃)

Measurement precision 

±0.2°C

Calibration error

±0.5%RH (0 … 90%RH)

Calibration error

±0.1°C (20°C)

±0.8%RH (40 … 95%RH)

τ63 response time

30s (probe fitted with stainless steel sintered filter)

- -

- -

Humidity sensitive element

High molecular polymer humidity-sensitive capacitor

Temperature sensitive element

Pt1000 (Grade 1/3B, DIN EN 60751)

The above precisions are comprehensive precisions, including repeatability, linearity and hysteresis of the measurement, and can be traced to the CNAS standard.

Technical Specification

Wiring connection

The maximum cable diameter of the terminal is 1.5mm2

Cable connector

M16 gland connector

Shell material

Stainless steel + PC (flame-retardant grade UL94-V0)

Protection grade

 IP65

Probe protection

Stainless steel sintering, H2O2, PTFE, etc.

Operating temperature

-20 ... 80℃

Electromagnetic compatibility

EN61326-1  EN61326-2-3

Storage temperature

-20 ... 60℃


Data download:

TH120-E Clean Industrial Wall-Mounted Temperature and Humidity Transmitter Product Manual





Copyright © GemCreate Electronics Technology (Shanghai) Co., LTD. All Rights Reserved.
Complaint feedback
Welcome feedback, your suggestion is our motivation!
We will carefully review each question you have feedback and give you the first reply.