Technical Features
Output signal: two ways 4 ~ 20mA, RS485, relay switching quantity
Output parameter: temperature, relative humidity, dew point, frost point, wet bulb temperature, enthalpy value, partial pressure of water vapor, mixed ratio, absolute humidity, moisture in oil
Shell material: 316 stainless steel
Structural feature: the upper and lower shell use detachable hinge design
Type test: nuclear grade type test
Configuration software: CONG-01 visual upper computer configuration software
Self-checking function: probe heating self-checking
Probe pressure resistance: 20Bar, immersible
Probe structure: push-pull self-locking aviation plug
Irradiation dose:
transmitter 0.35kGy cumulative gamma irradiation dose, yellow zone
transmitter and cable 42kGy cumulative γ irradiation dose, red zone
Probe cable: up to 100m in length
Measurement technology: small capacitance signal processing under the background of long cable and strong noise
Typical Applications
High temperature steam leakage monitoring in nuclear island area of nuclear power stations
Nuclear waste dew point parameter monitoring
High-temperature steam leakage monitoring of nuclear power units
Temperature and humidity environmental monitoring of nuclear-powered ships
Technical Data
Input and output
Operating voltage
|
24V DC ±20% or 220V AC
|
|
Current output
|
4 ~ 20mA RL < 500 Ohm
|
Power consumption
|
General mode: 70mA
|
|
|
Channel precision: ±0.2%FS
|
|
Self-check mode: 80mA
|
|
Data output
|
RS485 (Modbus RTU)
|
Switching quantity parameter
|
Contact resistance ≤ 100mΩ( 1A 6VDC)
|
|
Switching quantity output
|
Electric shock load: 5A 250VAC or 5A 30VDC
|
|
Mechanical durability: 1×107 times
|
|
|
Maximum switching voltage: 30VDC/277VAC
|
|
Electrical durability: 1×105 times (room temperature 1.5s on-off)
|
|
|
Maximum switching current: 8A
|
Measurement Parameter
Relative humidity
|
|
|
Temperature
|
|
Measurement range
|
0 ~ 100%RH
|
|
Measurement range
|
−40 … 190°C
|
Measurement precision
|
±2.5%RH (0 … 90%RH & 20℃)
|
|
Measurement precision
|
±0.2°C (20℃)
|
Calibration error
|
±0.5%RH (0 … 40%RH)
|
|
Calibration error
|
±0.1°C
|
|
±0.8%RH (40 … 95%RH)
|
|
|
|
τ63 response time
|
30s (probe fitted with stainless steel sintered filter)
|
|
- -
|
- -
|
Humidity sensitive element
|
High molecular polymer humidity-sensitive capacitor
|
|
Temperature sensitive element
|
PT1000 (Grade 1/3B, DIN EN 60751)
|
Humidity measurement drift
|
<1%RH/year (neutral clean environment)
|
|
Temperature measurement drift
|
<0.05℃/year (neutral clean environment)
|
The above precisions are comprehensive precisions, including repeatability, linearity and hysteresis of the measurement, and can be traced to the CNAS standard.
Mechanical Specification
Wiring connection
|
The maximum cable diameter of the terminal is 1.5mm2
|
|
Cable connector
|
M16*1.5 or M20*1.5 gland connector
|
Shell material
|
316 stainless steel
|
|
Connector diameter
|
Φ4...8 or Φ6...12
|
Probe protection
|
Stainless steel sintering, stainless steel grille, stainless steel pore, etc.
|
|
Probe cable diameter
|
Φ6 (irradiation cable); Φ4 (high-temperature cable)
|
Protection grade
|
IP65
|
|
|
|
Operating Environment
Operating temperature
|
Transmitter: -40 ... 80℃; Probe: -40 ... 190℃
|
|
Storage temperature
|
-20 ... 60℃
|
Type Test
Environment test
|
Standard
|
|
Reference test
|
Standard
|
High temperature test
|
GB/T 2423.2-2008
|
|
Visual inspection
|
RCC-E MC2000
|
Low temperature test
|
GB/T 2423.1-2008
|
|
Front-end equipment performance test
|
Response time, tolerance acceptance
|
Temperature change test
|
GB/T 2423.22-2012
|
|
Electrical test
|
RCC-E MC3000
|
Alternating heat and humidity test
|
GB/T 2423.4-2008
|
|
Electromagnetic compatibility test
|
Standard
|
Mechanical vibration test
|
GB/T 2423.10-2008
|
|
|
IEC 61000 (GB/T 17626)
|
Long run test
|
EJ/T 1197-2007
|
|
|
IEC 62003
|
Seismic test
|
HAF-J0053 / GB13625
|
|
|
|
Irradiation test
|
Transmitter: 0.35kGy cumulative γ irradiation dose, yellow zone
|
|
|
|
|
Probe cable: 42kGy cumulative γ irradiation dose, red zone
|
|
|
|
Data download:
WGTH-310N Nuclear Industry K3 Level Harsh Environment (Red Zone) Temperature and Humidity Transmitter Product Manual